Catalog Number 328418 |
Device Problems
Break (1069); Detachment Of Device Component (1104)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/22/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported before use of the bd insulin syringe with the bd ultra-fine¿ needle the consumer reported the needle broke off in the vial while he was drawing insulin.He does not re-use syringes.There was no report of injury or medical intervention.
|
|
Manufacturer Narrative
|
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch# 5313927.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Severity: s_1__; occurrence: a complaint history check was performed and this is the 1st related complaint for needle breaks off in vial on lot # 5313927.Based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Based on the above, no additional investigation and no capa is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Search Alerts/Recalls
|