Model Number 030424L
Medical Device Problem Codes Component Missing (2306); Mechanics Altered (2984)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the staples did not deploy for the reload was empty, it was without staples.The surgeon used another reload for the end of the procedure.No patient injury.
 
Additional Manufacturer Narrative
Additional information: evaluation summary post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted the reload displayed a full complement of staples.Damage was noted to the firing knob side of the knife blade channel.Microscopic inspection revealed no damage to the knife blade.Functional testing was precluded due to the observed condition of the device.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.Replication of this condition may occur if an excessive force is applied to the knife blade assembly prior to firing.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIA PREMIUM
Common Device Name
APPARATUS, SUTURING, STOMACH AND INTESTINAL
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7280732
Report Number2647580-2018-00830
Device Sequence Number1237099
Product Code FHM
UDI-Device Identifier20884523007098
UDI-Public20884523007098
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K801590
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model Number030424L
Device Catalogue Number030424L
Device Lot NumberP6L0622X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Initial Date Received by Manufacturer 02/05/2018
Supplement Date Received by Manufacturer05/11/2018
Initial Report FDA Received Date02/19/2018
Supplement Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-