BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE Back to Search Results
Catalog Number 320440
Device Problems Detachment Of Device Component (1104); Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 01/11/2018
Event Type  Injury  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd insulin syringe with the bd ultra-fine¿ needle separated and the needle remained in the skin.An x-ray was performed and the needle was removed from the skin with no surgical intervention required.
 
Manufacturer Narrative
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.If samples are received in the future the complaint will be reopened for further investigation.Unable to perform dhr check due to unknown lot number for needle break off in use.Severity: s_3__; occurrence: unable to perform complaint lot history check due to unknown lot number for needle break off in use.Based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Based on the above, no additional investigation and no capa is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE
Type of Device
INSULIN SYRINGE
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
MDR Report Key7280967
MDR Text Key100404718
Report Number1920898-2018-00049
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Catalogue Number320440
Device Lot NumberUNKNOWN
Initial Date Manufacturer Received 01/26/2018
Initial Date FDA Received02/19/2018
Supplement Dates Manufacturer Received01/26/2018
Supplement Dates FDA Received04/24/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-