| Model Number |
LF1744 |
| Medical Device Problem Code |
Energy Output To Patient Tissue Incorrect (1209)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the device had inadequate seal that lead to delayed bleeding of the operation site.The bleeding occurred during a single procedure and no re-operation was required.There was no alarms and no transfusion was required.No patient injury has been noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.Functional testing was performed with the returned device on porcine kidney tissue.Multiple seals on various size vessels were made and a full thermal effect was visually verified.The device was activated while pressing the button and with the rotation knob in various positions to detect any problematic activation issues.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|