| Model Number |
V068010 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Vascular Dissection (3160); Pericardial Effusion (3271)
|
| Date of Event |
01/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Conclusion not yet available, evaluation in process.This initial mdr is being submitted to meet our requirements of reporting.A follow-up will be submitted as soon as the investigation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported dissection of coronary sinus, slight pericardial effusion.During lao 30 degree angulation with balloon directed coronary sinus angiography dissection of the coronary sinus with visible slight pericardial effusion.Withdrawal of the hook sheeth.The right atrial and right ventricular lead were left in place, the device implanted and connected to the crt-device.Programming aai 30 0,2v/0,1ms.Tte verification of non-significant pericardial effusion.The patient was transferred hemodynamically stable to the intensive care unit for monitoring.Re-do procedure with left ventricular lead placement is planned in 4 weeks.We wanted to represent the cs from two levels.At the first presentation with balloon in the catheter in the ap, there was no visible perforation of the cs.To get a better overview, we wanted a second just view.We then set lao to 30 degree.And in this presentation of the cs saw a perforation of the cs.Has a possible deficiency with the device been observed? no.Did the event occur during surgery? yes.Did the event cause an adverse change in the patient's condition? yes dissection of coronary sinus, slight pericardial effusion.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of device history record (dhr) was conducted and there were no manufacturing rejects or anomalies recorded in the dhr affecting released product.The released device passed all in-process and qa final inspection steps before shipping to the customer.A complaint review of the reported lot found no additional reports involving this lot number.The device was not returned for analysis.There was no deficiency with the device observed during the procedure and no specific failure reported to oscor involving the device.This event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type.
|
| |
|
Search Alerts/Recalls
|