Medtronic received report of pipeline flex pushwire break during a procedure.The patient was undergoing flow diversion treatment for an unruptured, saccular aneurysm in the left internal carotid artery (ica).Vessel tortuosity was described as normal.The aneurysm had a max.Diameter of 3.29mm and neck diameter of 3mm.The landing zone artery size was 3.4mm distal and 4.5mm proximal.The devices were prepared as indicated in the ifu.During the procedure, the pipeline flex was reportedly deployed without difficulty.The physician reportedly saw sub-optimal wall apposition in the proximal section of the pipeline flex, but went ahead with deployment.At recapture of the pipeline flex delivery system, the physician attempted to advance the catheter over the pushwire, but was not able to.Instead, the catheter and pipeline flex pushwire were withdrawn from the patient together.After removal, it was observed that the distal section of the pushwire remained in the artery.Another physician arrived who used a balloon to angioplasty the proximal section of the pipeline flex open.Afterward, the physician used a gooseneck snare to capture the pushwire piece that had detached.The angiographic result post-procedure was good.The patient is reportedly fine.
|