| Model Number |
ZKB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Nausea (1970); Dizziness (2194); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown/not provided.Date of event - exact date of event is unknown, best estimate is between 2017- (b)(6) 2018.If implanted; give date(s): unknown, not provided.If explanted; give date(s): unknown, not provided (b)(4).An attempt has been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zkb00 intraocular lens (iol), 18.5 diopter was explanted from the right eye (od) of a female patient due to constant flickers of light in the eye.The patient complained of having trouble with balance and nausea and had blurry vision in the right eye after cataract surgery.The patient reports of decrease in flickers when she wears sunglasses or when is in a dark room.A non-amo lens was used as the replacement lens.There was no incision enlargement, no vitrectomy, and no sutures required at explant.There was no patient post-op injury reported.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received therefore the following fields were updated accordingly: age/date of birth: (b)(6).If explanted: give date(s): (b)(6) 2017.In addition it was learned the affected eye was the right eye.Initial reporter name and title: doctor (b)(6).Health professional: yes.Occupation: physician.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 03/15/2018.Device returned to manufacturer? yes.Device evaluation: the product was received in a pouch that was inside of an zip lock bag.Visual inspection was performed by a qualified operator using 12x magnification shows the product is cut in half, most probably to make explant possible.It also shows a cut and a scratch on the posterior side.The lens is contaminated with dust particles.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile or non-environmental controlled area.Investigation of the return sample and the condition of the return sample did not suggest the scratch was introduced during manufacturing.The reported complaint cannot be confirmed.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|