STRYKER SPINE-US ES2 90 DEGREES ROD INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 48280090
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/22/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an inserter would not disengage from rod.Surgeon disengaged lock screw and it would not release as it was hung up on the hex.A cobb was used to push it off an it would not release.Surgeon had to take the rod back out and increase the incision.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Method: visual inspection, device history review, complaint history review, risk assessment.Result: the customer reported event was confirmed via communication and visual inspection.Upon visual inspection, it was confirmed that the opening where the rod is inserted is deformed.Manufacturing history was reviewed and no issues were identified.The actual event led to increase in incision.A medical expert reviewed this complaint and concluded that increase in the incision does not poses serious harm to the patient.No surgical delay or adverse consequences to the patient is reported.Conclusion: based on observed wear and tear marks upon visual inspection and the age of the device (4 years), the plausible root causes of this event is normal wear and tear of instruments.
 
Event or Problem Description
It was reported that an inserter would not disengage from rod.Surgeon disengaged lock screw and it would not release as it was hung up on the hex.A cobb was used to push it off an it would not release.Surgeon had to take the rod back out and increase the incision.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ES2 90 DEGREES ROD INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7281700
Report Number0009617544-2018-00047
Device Sequence Number14028566
Product Code LXH
UDI-Device Identifier07613327006964
UDI-Public(01)07613327006964
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number48280090
Device Catalogue Number48280090
Device Lot Number138575
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Initial Date Received by Manufacturer 01/22/2018
Supplement Date Received by Manufacturer04/26/2018
Initial Report FDA Received Date02/20/2018
Supplement Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight84
-
-