MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 CORAIL2 NON COL HO SIZE 16; HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 CORAIL2 NON COL HO SIZE 16; HIP FEMORAL STEM Back to Search Results
Catalog Number L20316
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Codes Necrosis (1971); Pain (1994); Tissue Damage (2104); Discomfort (2330); Limited Mobility Of The Implanted Joint (2671)
Date of Event 11/30/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Alleges elevated chromium and cobalt levels in blood and mri results showed necrosis at the left hip joint.It was also alleged that there was loose hip, pain, metallosis, discomfort, injuries, reduced mobility, and emotional distress.After review of medical records for mdr reportability, patient was revised to address failed left metal on metal total metallosis and obesity.Operative notes reported of significant metallosis, pseudocapsular formation, significant trunnionosis/ corrosion, blackening of the trunnion, and extensive black corrosive appearing material behind the metal liner.Laboratory result for metal ions were below 7.Chromium: 1.6 pbb.Cobalt: 3.2 pbb.Doi: (b)(6) 2009; dor: (b)(6) 2015; left hip ((b)(6)).
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 NON COL HO SIZE 16
Common Device Name
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key7281821
Report Number1818910-2018-53683
Device Sequence Number19288629
Product Code KWL
UDI-Device Identifier10603295258070
UDI-Public10603295258070
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K123991
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2017
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date06/22/2011
Device Catalogue NumberL20316
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/29/2018
Supplement Date Received by Manufacturer11/08/2018
Initial Report FDA Received Date02/20/2018
Supplement Report FDA Received Date11/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient Weight159
-
-