| Model Number |
166818 |
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
01/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the user could not keep the balloon inflated during a period of two days.The balloon lost 30cc in that time.
|
| |
|
Event or Problem Description
|
|
It was reported that the user could not keep the balloon inflated during a period of two days.The balloon lost 30cc in that time.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed as cause unknown.Received 1 used silicone catheter for evaluation.Per visual evaluation, no defects along the catheter were found.During functional evaluation, the balloon was inflated with 15cc of air, using a syringe, and it was deflated.The catheter balloon was then inflated with 10ml of a mix of tap water and blue methylene using a syringe and it was deflated due to two pin-holes in the balloon.The balloon active length was found to be within specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 30cc balloon: use 35ml sterile water do not exceed recommended capacities.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use" 1149, 2199: "nl".
|
| |
|
Search Alerts/Recalls
|