Clinical evaluation performed by medical affairs on 20 february 2018: insert revision in tka 5 months after primary for joint instability.The insert was exchanged to a thicker one.The developmental instability is a rather commonly described possible complication following tka, because soft tissue may stretch and provide insufficient stability.It is therefore common practice to resort to a thicker insert and recreate soft tissue tension.This is not a problem due to any implant defect or malfunction.Batch review performed on 20 february 2018: lot 154782: (b)(4) items manufactured and released on 20 january 2016.Expiration date: 2021-01-03.No anomalies found related to the issue.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|