• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE; GUIDE WIRES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE; GUIDE WIRES Back to Search Results
Catalog Number 1001780JS
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.The other bmw guide wire referenced is being filed under a separate medwatch report.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified left anterior descending artery.It was noted that during an attempt to shape the tip of a balance middleweight 014 (bmw) guide wire (outside of the anatomy), the tip was noted to be peeling.Therefore, the device was replaced with another bmw 014 guide wire; but peeling of the tip was also noted on this device.Another unspecified bmw guide wire was then used to successfully complete the procedure.There was no patient involvement.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE
Common Device Name
GUIDE WIRES
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7282482
Report Number2024168-2018-01234
Device Sequence Number19289337
Product Code DQX
UDI-Device Identifier08717648025976
UDI-Public08717648025976
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K101116
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Catalogue Number1001780JS
Device Lot Number7102471
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2018
Initial Report FDA Received Date02/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-