MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 6.5MM LOW PROFILE HEX SCREW 25MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 6.5MM LOW PROFILE HEX SCREW 25MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 7030-6525
Medical Device Problem Codes Metal Shedding Debris (1804); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Primary procedure, right hip.It was reported that the screw passed entirely through the cup.The screw could not be extracted and was left in the patient but a metal shaving was retrieved from the cup.The surgeon placed another screw in the cup to provide fixation.No surgical delay was caused.
 
Additional Manufacturer Narrative
An event regarding an assembly issue and crack/fracture involving a bone screw was reported.The event was confirmed following a review by a clinical consultant.Conclusions: a review by a clinical consultant concluded: [.] implantation of a hex bone screw intended for supplemental screw fixation under a rather steep angle caused the relatively sharp edge of the flat screw head to cut through the relatively soft titanium of the cup shell perforating same.No adverse consequences are to be expected now or in the future.If additional information and/or device become available, this investigation will be reopened.
 
Event or Problem Description
Primary procedure, right hip.It was reported that the screw passed entirely through the cup.The screw could not be extracted and was left in the patient but a metal shaving was retrieved from the cup.The surgeon placed another screw in the cup to provide fixation.No surgical delay was caused.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5MM LOW PROFILE HEX SCREW 25MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7282608
Report Number0002249697-2018-00488
Device Sequence Number13755607
Product Code LPH
UDI-Device Identifier07613327345001
UDI-Public07613327345001
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K161569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Catalogue Number7030-6525
Device Lot NumberAYUA
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/23/2018
Supplement Date Received by Manufacturer03/12/2018
Initial Report FDA Received Date02/20/2018
Supplement Report FDA Received Date04/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
-
-