• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIORAPTOR 2.9 WITH 2 ULTRABRAID; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIORAPTOR 2.9 WITH 2 ULTRABRAID; SCREW, FIXATION, BONE Back to Search Results
Model Number 72200775
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/06/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a bancart and slap repairing, after drilling field with 2,7 mm drill device leaved from the grip.Patient bone was hard.Anchor was broken and all pieces were removed.Operation was completed with a titanium anchor by preparing new hole.
 
Additional Manufacturer Narrative
One 2.9 bioraptor ab assembly was returned for evaluation.Visual assessment of the device showed the anchor is missing from the inserter and was not returned for evaluation.The suture remains attached to the inserter indicating the anchor broke at its suture eyelet.Without the return of the anchor no root cause can be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIORAPTOR 2.9 WITH 2 ULTRABRAID
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key7282937
Report Number1219602-2018-00242
Device Sequence Number11242167
Product Code HWC
UDI-Device Identifier03596010585967
UDI-Public(01)03596010585967(17)180831(10)50563852
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model Number72200775
Device Catalogue Number72200775
Device Lot Number50563852
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2018
Initial Date Received by Manufacturer 02/06/2018
Supplement Date Received by Manufacturer03/22/2018
Initial Report FDA Received Date02/20/2018
Supplement Report FDA Received Date03/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age25 YR
-
-