Device is a combination product.(b)(4) device evaluated by mfr.: promus premier mr, ous, 3.5x38mm, stent delivery system was returned for analysis.A visual examination of the crimped stent found that the mid to distal end of the crimped stent was stretched and damaged.As a result the distal end of the stent was pulled out beyond the tip of the catheter.The undamaged stent outer diameter was measured which is within the maximum crimped stent specification.As part of the examination, the balloon was inflated to its rated burst pressure with no restrictions noted.No issues were observed with the balloon wall.The undamaged section of the stent (that was positioned over the balloon) deployed with no issue.It was not possible to deploy the stent for its entire length as the distal end of the stent was damaged and pulled over the tip and not placed over the balloon.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination found a hypotube kink 9.2 cm distal to distal end of strain relief.This most likely occurred during handling of the device during the procedure.A visual and tactile examination found no issue with the extrusion.The bi-component bond showed no signs of damage or strain.A visual and tactile examination found damage to the tip.The type of damage evident is consistent with resistance being encountered during advancement.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
Reportable based on analysis completed on 31-jan-2018.It was reported that the stent failed to deploy.The target lesion was located in a coronary artery.A 38x3.50mm promus premier¿ drug-eluting stent was advanced for treatment.However, it was noted that stent could not be deployed.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed stent damage.
|