Model Number 174233
Device Problem Product Quality Problem (1506)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter: occurred during a laparoscopic cholecystectomy procedure.The devices were being used for the dissection.Both devices had a burn at the end of the roticulator.Both devies were able to grasp properly and the handles did not bind.There was no patient harm.The patient outcome has no injury.
 
Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.Pmv found shearing damage of the plastic cable sheath.Engineering confirmed damage and found to be consistent with a mishandling of the device.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the reported condition may occur where an excessive force or leverage is applied to the device while retracting the clevis assembly and the clevis boot is subsequently damaged.This subsequent damage may result in improper retraction and deployment of the jaws during articulation and may cause shearing of the plastic cable sheath.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GRASP
Type of Device
MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7283154
MDR Text Key100561262
Report Number2647580-2018-00856
Device Sequence Number1
Product Code GET
UDI-Device Identifier20884523000815
UDI-Public20884523000815
Combination Product (y/n)N
Reporter Country CodeID
PMA/PMN Number
K914753
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model Number174233
Device Catalogue Number174233
Device Lot NumberP7C0694X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age38 YR
Patient Weight63
-
-