| Model Number |
E SERIES |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib fault 108" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The complainant was contacted for return of the device.The device has not been returned to zoll for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed and attributed to a system interconnect flex cable that was not properly seated.The cable and interlock were replaced to correct the reported problem.Analysis of reports of this type has not identified an increase in trend.
|
| |
|