| Catalog Number |
2N8378 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Clear passage testing and functional testing were performed with no issues noted.A pressure test was performed and revealed a leak from between the luer activated valve and the winged female luer lock adapter.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
During evaluation of a returned clearlink catheter extension set, the set was found to leak from between the luer activated valve and the winged female luer lock adapter.No additional information is available.
|
| |
|