Siemens healthcare diagnostics has determined that the cause of the discordant patient results on four separate patient samples observed on the bcs xp system (serial number (b)(4)) between the three (3) prothrombin time (pt) assays (pt, ptx (extended clot time), pt570) was due to a loose sample syringe plunger connector.The siemens field service engineer replaced the sample probe and sample tubing.Validation and quality control were run with acceptable recovery.The system was operational upon departure.The instrument and reagent are performing according to specifications.No further evaluation of this device is required.
|
Discordant patient results on four separate patient samples were observed on a bcs xp system (serial number (b)(4)) between the three (3) prothrombin time (pt) assays (pt, ptx (extended clot time), pt570).None of these results were reported to the physician.Pt quality control recovered within range.The same patient samples were repeated on an alternate bcs xp system (serial number (b)(4)) and generated pt patient results that matched the pt570 patient results generated on the bcs xp (serial number (b)(4)).Pt patient results were reported from the alternate bcs xp system (serial number (b)(4)).No erroneous results were reported.There are no reports of patient intervention or adverse health consequences due to the discordant pt patient results generated on the bcs xp system (serial number (b)(4)).
|