| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
Dizziness (2194)
|
| Date of Event |
01/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause : mlc-55 user had an inaccurate reference: competitor's meter: the end user is comparing results obtained from trividia's bgm system to the results from a competitor's bgm system.Test strip (b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with meter to meter comparison of 68 mg/dl using true metrix meter vs 51 mg/dl using another meter and a results of 198 mg/dl.The expected fasting blood glucose test result range is 98 mg/dl.The customer did report symptoms of feeling hungry and dizzy upon standing.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 04/30/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).Customer states that the doctor change his insulin from levemir to trisabea on (b)(6) 2018.Customer is now feeling hungry, and if he stands up, he feels dizzy.
|
| |
|