| Model Number |
CSK-TC15 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Expiration date: na.Additional suspect medical device component involved in the event: model: csk-tc10 lot: l884 description: csk tc electrode, 10cm.
|
| |
|
Event or Problem Description
|
|
A report was received that the wire electrode was broken off from the hub of the csk-tc15, and that the plastic covering at the hub was cracked on the csk-tc10 electrode.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to fu 1 report in field.It should have been (b)(6) 2018.
|
| |
|
Event or Problem Description
|
|
A report was received that the wire electrode was broken off from the hub of the csk-tc15, and that the plastic covering at the hub was cracked on the csk-tc10 electrode.
|
| |
|
Additional Manufacturer Narrative
|
|
The device analysis performed on csk-tc15, s/n (b)(4) revealed that the shaft of the electrode was found to be completely separated from the hub due to excessive external mechanical force.The functional test could not be performed due to the separated shaft.The complaint of the plastic covering at the hub being cracked on the csk-tc10 s/n (b)(4) was confirmed.Visual inspection found that the blue insulation that covers the hub of the electrode was damaged.However, the electrode passed the functional test.
|
| |
|
Event or Problem Description
|
|
A report was received that the wire electrode was broken off from the hub of the csk-tc15, and that the plastic covering at the hub was cracked on the csk-tc10 electrode.
|
| |
|