| Model Number |
M001262260 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported there was difficulty deploying the array.A 2.0/17/15 leveen¿ superslim¿ was selected for a radiofrequency ablation procedure.When attempting to deploy the device while outside of the patient, the handle portion was difficult and it was unable to be deployed smoothly.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer, evaluation summary attached, method codes, result codes, conclusion codes: updated.Device evaluated by manufacturer: examination of the returned complaint device revealed that the electrode array was fully retracted.No visible damages were found to cannula or handle.A functional evaluation extended and retracted the tines without resistance.The tines were evenly spaced and undamaged.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported there was difficulty deploying the array.A 2.0/17/15 leveen¿ superslim¿ was selected for a radiofrequency ablation procedure.When attempting to deploy the device while outside of the patient, the handle portion was difficult and it was unable to be deployed smoothly.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|