| Model Number |
SA-85 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant iis stent graft system was implanted in a patient for the endovascular treatment of an 7.2cm abdominal aortic aneurysm with approx 5mm of proximal neck length.It was reported that after implantation of the endurant bifurcate and the placement of 6 helifx endoanchors, both in good proximal position, a completion angiogram showed a type ia endoleak.As per the physician the cause of the event is anatomy related and was due to the short, conical proximal neck.It was reported that intervention was performed, diagnostic angiogram demonstrated approximately 1mm of proximal neck could be achieved by placing an aortic cuff.The physician made the decision to balloon the proximal stent graft several times, with the addition of 8 endoanchors implanted at the proximal end of the stent graft.A follow up angiogram showed a significant reduction of the endoleak after the intervention.Additional information received reported that during a follow up ct, no endoleak was observed.No clinical sequelae were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received reported that during a follow up ct, no endoleak was observed.If information is provided in the future, a supplemental report will be issued.
|
| |
|