• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1032800
Medical Device Problem Codes Incorrect Or Inadequate Test Results (2456); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/01/2018
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer for service.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, the device failed a test step during testing.The device's active exhalation control module was replaced to address the issue.
 
Additional Manufacturer Narrative
The manufacturer previously reported a ventilator's active exhalation control module was replaced due to a failed test step.During further evaluation the device's sensor board was replaced to address a "service required" alarm condition.The device's sensor board was returned to the manufacturer's quality assurance laboratory for further investigation.The root cause of the sensor board failure was due to the pressure sensors mt3 and mt2 being out of tolerance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
MDR Report Key7285213
Report Number2518422-2018-00380
Device Sequence Number11176590
Product Code CBK
UDI-Device Identifier00606959026360
UDI-Public00606959026360
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Service and Testing Personnel
Type of Report Initial,Followup
Report Date (Section B) 02/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1032800
Device Catalogue Number1032800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/01/2018
Supplement Date Received by Manufacturer05/04/2018
Initial Report FDA Received Date02/21/2018
Supplement Report FDA Received Date05/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-