MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA MPACT DOUBLE MOBILITY ACETABULAR SHELL Ø48

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA MPACT DOUBLE MOBILITY ACETABULAR SHELL Ø48 Back to Search Results
Catalog Number 01.32.148MB
Health Effect - Clinical Code Joint Disorder (2373)
Date of Event 01/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 19 february 2018 lot 155035: (b)(4) items manufactured and released on 12 october 2015.Expiration date: 2018-02-28.No anomalies found related to the problem.To date, all items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medical affairs: partial hip revision surgery (cup, head and liner) occurred 2 years after primary total hip arthroplasty.Radiographic image provided shows the presence of a horizontal cup.The reason of this position cannot be assessed having a single pre-revision low-quality x-ray.The cause of this event cannot be determined on the basis of available information.
 
Event or Problem Description
The patient came in complaining of iliopsoas pain due to cup mobilization.The surgeon revised the cup, head and liner.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MPACT DOUBLE MOBILITY ACETABULAR SHELL Ø48
Common Device Name
ACETABULAR SHELL
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7285365
Report Number3005180920-2018-00072
Device Sequence Number1993924
Product Code LPH
UDI-Device Identifier07630030860959
UDI-Public07630030860959
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143453
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Catalogue Number01.32.148MB
Device Lot Number155035
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2018
Initial Report FDA Received Date02/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
-
-