• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PINN MAR NEUT 28IDX46OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PINN MAR NEUT 28IDX46OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121928046
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Implant had to be revised due to luxation between cup and liner.Doi: unknown ; dor: (b)(6) 2018; unknown hip.
 
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: visual examination of the returned device finds evidence to support disassociation of the liner and cup while implanted.Device evaluation did not however identify or verify product error with regard to the reported event.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 28IDX46OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7285509
Report Number1818910-2018-53813
Device Sequence Number1824039
Product Code LPH
UDI-Device Identifier10603295013273
UDI-Public10603295013273
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K033273
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number121928046
Device Lot NumberH13527
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2018
Initial Date Received by Manufacturer 02/02/2018
Supplement Date Received by Manufacturer09/17/2018
11/19/2018
Initial Report FDA Received Date02/21/2018
Supplement Report FDA Received Date09/20/2018
11/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-