Catalog Number L20409 |
Device Problem
Scratched Material (3020)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that when the surgeon was ready to trial components, the corail standard neck trial would not snap onto the size 9 corail trial stem.After further observation of the size 9 trial stem, it showed increased wear and damage to the post on the stem trial where the neck trial slides over.All pieces were retrieved from the patient and no loosening of any component was indicated.No surgical delay.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).Investigation summary: the instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|