| Model Number |
ENVEOR-N-US |
| Medical Device Problem Codes |
Difficult to Remove (1528); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned for analysis, however, the analysis has not yet been completed.Conclusion: without the completion of product analysis, no definitive conclusions could be drawn regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the implant of a transcatheter bioprosthetic valve, the valve was recaptured.During reca pture, resistance was met while turning the delivery catheter system (dcs).The dcs was withdrawn from the body and a capsule separation was observed.A second valve was loaded to a second dcs and used for implant.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at medtronic's quality laboratory, the valve was received loaded in the capsule of the delivery catheter system (dcs).The handle appeared intact.The tip-retrieval mechanism appeared intact.Damage was observed on the threading of the screw gear.The device was returned with the end cap/screw gear snap fit connected.Voids were present on the capsule from the distal end to the proximal end.The proximal end of the capsule was observed to have separated.The separation site was jagged and uneven.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.The valve was recaptured twice, which is a feature of the enveor delivery system to allow re-positioning of the valve.The report indicates that resistance was noted during recapture, and the valve was withdrawn from the body, where a capsule separation was observed.Capsule separation typically occurs due to excessive compressive forces applied during the valve loading process.This excessive compressive force is typically experienced when the valve is loaded incorrectly.The subject dcs and valve loaded inside were received and evaluated by medtronic; the damages observed on the screw gear suggest excessive forces applied on the system during rotating the knob, and the observed capsule voids indicate delamination between the capsule outer polymer and the nitinol frame, and typically occur when the capsule is subjected to a bending force potentially after tracking through tortuous anatomy, or if a valve has been misloaded.In this case, there were no procedural cine/angio or load check images provided, therefore it cannot be independently determined if the valve was correctly loaded onto the dcs, and the series of events leading to the capsule separation could not be assessed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|