• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1080-060
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual, dimensional and functional inspections were performed on the returned device.The reported difficulty was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that contributed to this event.Additionally, a review of the complaint history revealed no other incidents.It is likely that the balloon folds and profile dimensions became compromised after the first inflation and removal.Based on the information reviewed and analysis of the returned device, the reported difficulties appear to be related to case circumstances.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported the procedure was to treat a lesion in the heavily calcified right common iliac.The 8mm x 60mm armada 35 balloon dilatation catheter (bdc) was advanced without issue for pre-dilatation.After pre-dilation was performed during removal, resistance was felt with the introducer sheath.The bdc and introducer sheath were removed as a single device.There were no adverse patient effects and there was no clinically significant delay during the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7286795
Report Number2024168-2018-01286
Device Sequence Number2273893
Product Code LIT
UDI-Device Identifier08717648154874
UDI-Public08717648154874
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Catalogue NumberB1080-060
Device Lot Number70613G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2018
Initial Report FDA Received Date02/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-