The customer reported that, during placement of the ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter in the patient's abdomen as part of a sclerotherapy procedure, the trocar needle wasn¿t protruding out of the end as expected.A direct stick was not possible, and the operator had to back the product out, which resulted in an inability to remove the needle from the catheter once it was in the patient.The device had to be removed, and a re-sticking of the patient was necessary.The device was reportedly only in place for a period of minutes, and the customer reported that no patient adverse events occurred as a result of the issue.The device has been returned for evaluation; however, as of the date of this report, the investigation is still pending.
|
|
A review of the complaint history, drawings, dimensional verification, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, trends, quality control, and visual inspection of the returned device was conducted during the investigation.The device was returned and the complaint has been confirmed per product examination.Device failure analysis was conducted and the stiffening cannula was measured and found to be within specifications.However, the examination verified that the catheter is too long.The cause has been determined to be manufacturer related.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|