To date, the device has not been returned yet, therefore the device has not been examined.The device was either lot number 171421 or 170934.Since the spiral was reported that it broke during insertion, it could have been from being excessively torqued.If the electrode is excessively torqued during (e.G.Rotated > 1 ½ turns clockwise) application to the presenting part, then this could have caused the needle to break and remain embedded in the patient's scalp.Based on the information reported, i feel that using excessive torque when applying the electrode to the baby's head is the cause of the breakage of the spiral needle tip.More education on proper insertion along with reviewing of the ifu's are indicated.
|