| Model Number |
MT22608-BLU |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Pt mother reports loss of connection.Allegies that the receiver is at fault.
|
| |
|
Additional Manufacturer Narrative
|
|
Com-(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4)2018 that on (b)(6)2018, a loss of connection between the transmitter and display device occurred.No product or data was provided for evaluation.The complaint confirmation of a loss of connection could not be determined.A probable cause could not be determined.No additional patient or event information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An external visual inspection was performed and passed.Charge and boot testing was performed and passed.A (b)(4) bluetooth device pairing test was performed and failed.A receiver functional test was performed and failed due to failure of the radio frequency (rf) test.A review of the share logs was not performed as there was no data found within the investigation window.The allegation was confirmed.No injury or medical intervention was reported.
|
| |
|