| Model Number |
A902416 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the iodine packet had burst in the foley tray.
|
| |
|
Event or Problem Description
|
|
It was reported that the iodine packet had burst in the foley tray.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed.Received 1 opened surestep foley tray system for evaluation.During the visual inspection, it was noted that the iodine package had a small cut in one side.The damage measured 0.136".The package presented a leak by this small cut and contaminated the tray.The root cause of the damage could not be determined.The damage could have occurred during handling or transportation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use." 1267: "l" 2645: "nl".
|
| |
|