Catalog Number 2968255
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent oblique lumbar interbody fusion (olif) procedure due to lumbar degeneration.Intra-op, the doctor found that the implanted cage was fractured.The cage was replaced with a new one.No patient complications were reported as a result of this event.No fragments of the broken product remained inside the patient.
 
Additional Manufacturer Narrative
Product analysis result: visual microscopic the cage was returned cracked into multiple pieces.A microscopic review of the fracture face is consistent with material overload during impaction.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLYDESDALE SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf 94469
MDR Report Key7288244
Report Number1030489-2018-00258
Device Sequence Number2379247
Product Code MAX
UDI-Device Identifier00613994485793
UDI-Public00613994485793
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P000058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2025
Device Catalogue Number2968255
Device Lot Number12ES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2018
Initial Date Received by Manufacturer 01/23/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
-
-