| Model Number |
MSERIES CCT |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device did not respond to the front panel controls.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing which included environmental and functional testing without duplicating the malfunction.The device's control board and an interconnection flex cable were replaced and the display cable ferrite filter was re-secured as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|