| Model Number |
0037860 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the reported device has not been returned to conmed for evaluation.Should the device be returned, this complaint will be reassessed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of this product.The products released for distribution were found to have met all specifications prior to shipment.A two-year historical complaint review revealed (b)(4) prior complaints for this device family and failure mode.In this same timeframe, (b)(4) units have been manufactured and shipped worldwide, making the rate of occurrence of this failure (b)(4) percent.A risk analysis was performed and deemed acceptable.This reported packaging issue was obvious to the distributor, prompting the return of the device for evaluation.There was no patient involvement.As with all medical devices, examination of the product occurs multiple times prior to use.Good clinical practice would include examination and verification of the original packaging and its labeling to ensure both are intact.The instructions for use (ifu) provides the following warning.If packaging has been opened/damaged or altered, do not use the product and contact the manufacturer immediately.This incident type will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The distributor in (b)(6) rejected one 0037860 surgical tubing due to an "insufficient heatseal".In this instance, there was no patient involvement as the packaging anomaly was discovered during inspection prior to distribution to an end-user.The report is raised on the basis of a device malfunction with potential for injury with recurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported unused devices were returned for evaluation in original unopened packaging.Label verification and visual inspection was performed.The device packaging's long straight seal was observed to have an open seal.Seal transfer was not observed within the defect area.Due to the appearance and location of the defect, it is likely that the defect was caused by improper device placement during the sealing process.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Search Alerts/Recalls
|