AV-TEMECULA-CT ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1013465-200
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history records revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lots revealed no other incidents.The investigation was unable to determine a cause for the reported difficulties.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a non-tortuous, moderately calcified anterior tibial artery.A 3.0 x 200mm armada 18 balloon dilatation catheter (bdc) was inflated once at 8 atmospheres.Resistance was met during removal.Once removed it was noted that part of the balloon was dangling off the distal shaft, but still attached.Another same size armada 18 bdc was used to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7289033
Report Number2024168-2018-01304
Device Sequence Number13788797
Product Code LIT
UDI-Device Identifier08717648207938
UDI-Public08717648207938
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number1013465-200
Device Lot Number7092541
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2018
Initial Report FDA Received Date02/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age84 YR
Patient Weight52
-
-