WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM TI CORTEX POLYAXIAL SHAFT SCREW 30MM F/4.0MM RODS; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
|
Back to Search Results |
|
| Model Number |
04.615.330 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Pain (1994); Nonunion/Delayed-union Bone Fracture (2369)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient information is unknown.Date of event is unknown.Implant date is unknown and occurred in 2013.Explant date is unknown.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Therapy date of concomitant device is unknown.The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history record review was requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a c1-c6 synapse 4.0 mm rod construct presented to the surgeon with pain, and was found to have a nonunion at c1-c2 and that the right c1 locking screw (04.614.508) has come completely out of the right c1 screw (unknown size and part #).The rod part # is also unknown.The hardware was possibly originally implanted on an unknown date in 2013.The surgeon plans to perform a revision on the patient in the near future.He plans to take out the c1 screws and cut the rods just cranial to the c2 screws.The surgeon does not intend to replace the removed hardware with other instrumentation.There are six devices involved in this complaint.The revision for the complaint reported previously took place (b)(6) 2018.The patient had a previous surgery where synapse 4.0 mm rod system was used on a posterior cervical.The construct was from c1-c5 with synapse 4.0 (no screws at c2).The patient never healed at the c1-c2 level and the surgeon determined that the patient has a nonunion there.At an unknown point in time, the hardware at c1 failed.On the left, the c1 screw broke off into the patients lateral mass, just distal to the synapse shaft on the shaft screw (serial (b)(4)).The locking cap was still threaded onto the screw securing it to the rod.On the right synapse c1 screw (04.615.330 serial (b)(4)) the locking cap unthreaded from the synapse screw and was free floating.Additionally, the head on this synapse screw popped off of the actual bone screw.All of these failures happened at an unknown time.During the revision, the surgeon removed both of the synapse locking caps from the patient.He then cut the rods in situ cranial to the c3 screws.He then removed the c1 screws, but did not remove the tip of the left c1 screw which remains in the patient.The patient has a solid fusion from c3-c5 and no hardware was placed at c1 and c2.Concomitant parts: qty 2, unk - rods: synapse.Qty 1, ti locking screw (part 04.614.508, lot unknown).This is report 2 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.Device is a single use device but was not reprocessed or reused.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
The subject device has been received by the manufacturer and is undergoing investigation.The results of the investigation are pending completion and will be submitted in a supplemental report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed for part number: 04.615.330, lot number: 7078542: manufacturing site: (b)(4), release to warehouse date: 06-nov-2012, supplier dhr: n/a: no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device condition: the polyaxial screw was received with a transverse break in the threaded shaft.The break is located approximately 3mm from the proximal start of the threaded region.The distal portion of the threads were not returned.The fracture site was examined and no material voids or irregularities were identified.The balance of the device shows surface wear consistent with implant in 2013 and subsequent removal in 2018 and which would not impact the functionality.Thus, the complaint condition is confirmed and consistent with the reported condition.Replication of the complaint condition is not applicable as the device is already broken.Lot number review: record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The lot went through the required steps during the inspection at the time of manufacturing and the dhr records showed no issues concerning the material.Release to warehouse date: 06-nov-2012 drawing review: based on the date of manufacture the relevant drawings, reflecting the current and manufactured revision, were reviewed.The minor thread diameter at the fracture site was measured and is within the specification per relevant drawings.The design history was found to not impact the complaint condition.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.No definitive root cause was able to be determined as circumstances surrounding the event, the details of the nonunion, and the conditions over the implant duration (2013 to 2018) are unknown.Conclusion: a device history record (dhr) review, device inspection, and drawing review were performed as part of this investigation.The complaint condition is confirmed as the screw was received with a transverse break in the threaded shaft.There were no issues during the manufacture of the product that would contribute to this complaint condition.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Thus, there is no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Reported incorrectly on initial medwatch and should have been reported as device is not a reprocessed single use device.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|