| Model Number |
8450107525 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is available for return.A follow up report will be filed once the quality investigation is complete.Awaiting device return to manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the most probable root cause of bur breakage is as a result of a bearing failure at the head of the associated attachment (5407120950c sn (b)(4)).The quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during service conducted at the manufacturer a broken bur was found inside the attachment.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
|
| |
|