• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY FLEXTOME® CUTTING BALLOON®; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY FLEXTOME® CUTTING BALLOON®; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H749CBM3300100
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that a catheter shaft break occurred.A 10/3.00 flextome cutting balloon was selected for use.During preparation, when they pulled the flextome balloon out of the hoop to press it, it was noted that there was a kink in the shaft about 4-6 inches from the hub.The device was not used into the patient's body.When the balloon was put back in the hoop, it was further noted that the shaft completely broke in two.The procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.A visual and tactile examination identified a complete break in the hypotube of the device.A kink on the hypotube was also identified located approximately 60mm distal of the strain relief.This type of damage is consistent with excessive force being applied to the delivery system.A visual examination identified that the balloon was tightly folded indicating that the balloon had not been subjected to positive pressure.A visual and microscopic examination identified no damage to the tip, markerbands or blades of the device.All blades were present and fully bonded to the balloon material.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event or Problem Description
It was reported that a catheter shaft break occurred.A 10/3.00 flextome cutting balloon was selected for use.During preparation, when they pulled the flextome balloon out of the hoop to press it, it was noted that there was a kink in the shaft about 4-6 inches from the hub.The device was not used into the patient's body.When the balloon was put back in the hoop, it was further noted that the shaft completely broke in two.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME® CUTTING BALLOON®
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7289574
Report Number2134265-2018-01235
Device Sequence Number16176933
Product Code NWX
UDI-Device Identifier08714729808350
UDI-Public08714729808350
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/26/2020
Device Model NumberH749CBM3300100
Device Catalogue NumberCBM330010
Device Lot Number21432434
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2018
Initial Date Received by Manufacturer 01/31/2018
Supplement Date Received by Manufacturer03/05/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date03/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-