| Model Number |
HFD100 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Clinical specialists visited the customer site to diagnose the issue.They observed the or staff set up and pin the patient and reported that there were no errors in how the hfd was being used.After the case, they attempted to make the device fail and could not.After a thorough examination, they found no issues.
|
| |
|
Event or Problem Description
|
|
The hfd100 head fixation device was being utilized during a surgical case and at some point intraoperatively, the brainlab navigation became inaccurate.There was no evidence of the patient slipping in the skull pins.There were no injuries reported.
|
| |
|