• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493808012450
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a coronary vessel.A 4.5mmx12mm quantum¿ maverick¿ balloon catheter was advanced for dilation.However, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in a coronary vessel.A 4.5mmx12mm quantum¿ maverick¿ balloon catheter was advanced for dilation.However, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The balloon was loosely folded.The shaft is intact; there is no shaft separation.The outer shaft, inner shaft, balloon and tip were microscopically examined.The tip is damaged.There are numerous hypotube kinks.There is shaft damage at the exit notch that is consistent with damage seen with the use of a guidewire.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.There was no evidence of any material or manufacturing deficiencies contributing to the damage.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7290095
Report Number2134265-2018-01145
Device Sequence Number1741936
Product Code LOX
UDI-Device Identifier08714729392729
UDI-Public08714729392729
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model NumberH7493808012450
Device Catalogue Number38080-1245
Device Lot Number0020485215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2018
Initial Date Received by Manufacturer 01/30/2018
Supplement Date Received by Manufacturer02/23/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date03/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-