| Model Number |
H7493808012450 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in a coronary vessel.A 4.5mmx12mm quantum¿ maverick¿ balloon catheter was advanced for dilation.However, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in a coronary vessel.A 4.5mmx12mm quantum¿ maverick¿ balloon catheter was advanced for dilation.However, the shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The balloon was loosely folded.The shaft is intact; there is no shaft separation.The outer shaft, inner shaft, balloon and tip were microscopically examined.The tip is damaged.There are numerous hypotube kinks.There is shaft damage at the exit notch that is consistent with damage seen with the use of a guidewire.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.There was no evidence of any material or manufacturing deficiencies contributing to the damage.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|
Search Alerts/Recalls
|