• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5532-G-413
Medical Device Problem Codes Detachment Of Device Component (1104); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock between with no reported discrepancies.There have been no other similar events for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
As reported: "surgeon, during bilateral total knees, tried to lock the insert in to the tibial tray of the right knee, when it didn¿t lock he used a osteotome to work the insert back out and when he did he stated that the wire fell off.This happened a second time with another insert in the same knee.We did not have any other insert as we had used two.Surgeon decided to leave an insert trial in and come back on friday(b)(6) 2018 to put another insert in at that time.".
 
Event or Problem Description
As reported: "surgeon, during bilateral total knees, tried to lock the insert in to the tibial tray of the right knee, when it didn't lock he used a osteotome to work the insert back out and when he did he stated that the wire fell off.This happened a second time with another insert in the same knee.We did not have any other insert as we had used two.Surgeon decided to leave an insert trial in and come back on friday (b)(6) 2018 to put another insert in at that time.".
 
Additional Manufacturer Narrative
An event regarding a seating/locking issue involving a triathlon insert was reported.The event was not confirmed.Conclusion: the exact cause of the event could not be determined as insufficient information was provided.Further information such as pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.A trend analysis was conducted for the reported failure mode and concluded that seating/locking difficulties may arise from user-related issues associated with the preparation of the joint space prior to the attempted seating and with the technique required to correctly introduce the insert component to the baseplate.No further investigation for this event is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7290213
Report Number0002249697-2018-00529
Device Sequence Number1671227
Product Code MBH
UDI-Device Identifier07613327050721
UDI-Public07613327050721
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Catalogue Number5532-G-413
Device Lot Number7M6VKY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Initial Date Received by Manufacturer 01/25/2018
Supplement Date Received by Manufacturer03/26/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date04/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age39 YR
-
-