Catalog Number 328508
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: n/a.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while drawing up insulin the needle of a bd¿ syringe detached from the syringe and remained in the medication vial, at least twice.There was no report of exposure, injury or medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.As per manufacturing, a review of the device history record was completed for batch# 5229834.All inspections and challenges were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Investigation conclusion: based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD RELION¿ INSULIN SYRINGE
Common Device Name
INSULIN SYRINGE WITH NEEDLE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key7290775
Report Number1920898-2018-00063
Device Sequence Number19427839
Product Code FMF
UDI-Device Identifier00681131311700
UDI-Public00681131311700
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K024112
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 04/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number328508
Device Lot Number5229834
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/29/2018
Supplement Date Received by Manufacturer01/29/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date04/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-