Catalog Number 1012538-40
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.Reportedly, the absolute pro stent delivery system was advanced to the right cephalic vein.It should be noted that the absolute pro instruction for use (ifu) states: the absolute pro vascular self-expanding stent system is indicated for improving luminal diameter in patients with de novo or restenotic atherosclerotic lesions in the native common iliac artery and native external iliac artery.It is unknown if the ifu deviation contributed to the reported difficulty.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a cause for the reported inaccurate delivery.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported the procedure was performed to treat a lesion with no calcification in the right cephalic vein.A 10.0/40 mm absolute pro stent jumped distally during deployment.The stent was placed partially in the lesion and partially in healthy tissue.An additional non-abbott stent was used to treat the remaining part of the lesion.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7290891
Report Number2024168-2018-01333
Device Sequence Number11341349
Product Code NIO
UDI-Device Identifier08717648176289
UDI-Public08717648176289
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Catalogue Number1012538-40
Device Lot Number5081361
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/08/2018
Supplement Date Received by Manufacturer02/22/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date02/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
-
-