COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LS3112
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
One device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The reported condition was not confirmed.The device was recognized and activated normally when tested.No alarms occurred and an end tone was heard each time indicating a completed seal.An alternative and reportable issue was identified, where one of the snap pins had broken off and was not returned.Snap damage can be caused by the user improperly misaligning the snaps into the mating holes during assembly.This damage can also be caused by multiple assembly attempts.The investigation identified the root cause of the reported event to be attributed to user error.Refer to the product instructions for use (ifu) for additional information on the proper method of assembling the electrodes onto the reusable instrument.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, there was a sealing issue with the device.The device would activate and continue to seal but would not reach an end tone.An alarm would occur.There was no patient injury.Examination of the received sample found one of the snap pins from the jaws of the device had broken off and was not returned.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7291029
Report Number1717344-2018-00258
Device Sequence Number11104557
Product Code GEI
UDI-Device Identifier20884524002511
UDI-Public20884524002511
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K043273
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2022
Device Model NumberLS3112
Device Catalogue NumberLS3112
Device Lot Number70890155X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2017
Initial Date Received by Manufacturer 01/28/2018
Supplement Date Received by Manufacturer03/01/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date03/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age38 YR
-
-