• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT400
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens remains implanted.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zct400 20.5 diopter intraocular lens (iol) was noticed rotated from its initial 160 degrees placement to 4 degrees in the patient's left eye (os) at the 6-month visit with the doctor.The lens was implanted (b)(6) 2017.Reportedly, there was no patient complications or related injury and the outcome does not significantly interfere with activities of daily life for the patient.There are no plans at the moment to reposition the lens.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: visual inspection/product testing could not be performed as the product was not returned for evaluation.The reported complaint could not be verified.Manufacturing record review: manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this production order number to date.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7291308
Report Number9614546-2018-00178
Device Sequence Number2378566
Product Code HQL
UDI-Device Identifier05050474553927
UDI-Public(01)05050474553927(17)200725
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/25/2020
Device Model NumberZCT400
Device Catalogue NumberZCT400U205
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/26/2018
Supplement Date Received by Manufacturer03/12/2018
Initial Report FDA Received Date02/22/2018
Supplement Report FDA Received Date04/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age87 YR
-
-