| Catalog Number |
7570905 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with degenerative disc disease and underwent minimally invasive lumbar decompression and fusion at l3-s1.During surgery, the sleeves each ¿popped¿ off when surgeon was attempting to reduce rod into screw head.The sleeves appear bent.The product came in contact with the patient.No patient complications were reported as the result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual, functional and microscopic examination found that one side of the reducer¿s tip is bent outward.This type of damage is consistent with overload through bending while attached to the screw.There are witness marks on the location tab indicating the screw had slipped off.The observation are consistent with bend stress overload.If information is provided in the future, a supplemental report will be issued.
|
| |
|