| Catalog Number |
01.12.046 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/17/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on (b)(6) 2018: lot 112621: 15 items manufactured and released on (b)(6) 2011.Expiration date: (b)(6) 2016.No anomalies found related to the issue.To date, 13 items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018 the sales representative has been informed by the surgeon that a revision surgery was performed.At this time we do not know if the reason of the revision surgery is bone fracture or stem loosening.
|
| |
|
Additional Manufacturer Narrative
|
|
Clinical evaluation performed by medical affairs director: from the xray analysis it appears that the stem may have subsided at some stage (no postoperative xray is available), thus creating a significant leg length discrepancy and a slightly valgus stem.Apart from this, we do not see other major problems in this tha.
|
| |
|