Clinical evaluation performed by medical affairs director: few days after primary hybrid tha on an elderly lady, femoral stem revision is required.The reason is not specified in the report.The radiographic images show a varus-positioned stem that probably left no cement on the medial proximal surface, a low resection level and a very distal femoral plug.The distal cement is thoroughly detached from the bone, reasons unknown, possibly because of insufficient cleaning or pressurization.There a re no indications that a defective device led to this adverse event.
|